Transistor Characteristic Curve Tracer
System Block Diagram
A PCI interface card and BJ4822 operating software are installed on the computer; the computer controls the BJ4822 test host through the PCI bus; simultaneously reads data from the acquisition memory of the BJ4822 test host; and displays the characteristic curves of semiconductor devices through the BJ4822 operating software. Point marking testing is provided. Test devices can be connected to the test box via test adapters or directly via test leads; each source is connected to the test box panel; users can select different sources to inject into the various pins of the test device as needed.
Main Technical Specifications
(1) Collector Pulse Source (High Current Source):
Short-circuit current: 3kW—maximum short-circuit current 400A ;
300W—maximum short-circuit current 40A ;
Open-circuit voltage: 30V (+10%, -5%)
Adjustable range: continuously adjustable from 0% to 100%, resolution 0.5%
Pulse width: 300μs±10%
Repetition frequency: 3kW—1/4 power frequency; 300W—1/2 power frequency
(2) Collector Sweep Source (High Voltage Source):
Short-circuit current: 30W—40mA; 3W—4mA;
Open-circuit voltage: 3kV (+10%, -0%)
Adjustable range: continuously adjustable from 0% to 100%, resolution 0.1%
(3) Step Voltage Source:
Step amplitude range: 200mV/step—5V/step (in 1-2-5 sequence); 5 ranges in total; error of each range ≤±3%
Number of steps: continuously variable from 0 to 5 steps
(4) Step Current Pulse Source:
Step range: 1mA/step— 2A /step (in 1-2-5 sequence); 11 ranges in total; error of each range ≤±3%
Pulse width: 550μs±10%
Number of steps: continuously variable from 0 to 5 steps
(5) Step Current Source:
Step range: 1μA/step—2mA/step (in 1-2-5 sequence); 11 ranges in total; error of each range ≤±3%
Number of steps: continuously variable from 0 to 5 steps
(6) Horizontal Voltage Measurement:
(a) Vce voltage source setting (in 1-2-5 sequence):
3kW, 300W: 100mV/div—5V/div (in 1-2-5 sequence); 5 ranges in total; error of each range ≤±3%
30W, 3W: 50V/div—500V/div (in 1-2-5 sequence); 4 ranges in total; error of each range ≤±5%
(b) Vbe voltage source
100mV/div—5V/div (in 1-2-5 sequence); 5 ranges in total; error of each range ≤±3%
(7) Collector Current Measurement:
Vertical range (in 1-2-5 sequence):