Desktop|中文
Home About Us Products + News Center Technical Articles Messages Contact Us

Product Details

全自动Dielectric Loss Tester-扬州达瑞

Current Transformer On-site Calibration Instrument - Yangzhou Dary

Published: 2020-08-21 09:46:28

Views: 2922

Category: CT/PT Characteristic Tester

Product Description
Current Transformer On-site Calibration Instrument



Dary Electric recommends to you



Main technical indicators:

>Current transformer error measurement part

①Overall accuracy: 1/3 of the error limit of the tested current transformer

②Test range: 5A/5A~25000A/5A or 5A/1A~6300A/1A  

③Working range of tested current transformer: 1% to 200% 

④Secondary load: 2.5VA~300VA, COSφ=0.1~1.0 

⑤Accuracy range of tested current transformer: 1.0, 0.5, 0.5S, 0.2 and 0.2S 

⑥Resistance and admittance measurement error ≤5.0%

>Measurement range: R: 0.00Ω~20.0Ω

       Y:  0.000mS~100.0mS

>Instrument power consumption: 20VA 

>Instrument accuracy class: 0.05S

>Maximum dimensions (cm): 460×375×183

>Weight (kg): 10



Current Transformer On-site Calibration Instrument has a built-in standard voltage transformer equivalent to the same ratio as the tested current transformer, with accuracy up to 0.05 class, accurately measuring the ratio and no-load error of the tested current transformer. Then, combined with the test results of impedance and admittance, the error of the transformer is calculated.

>Adopts near-power-frequency heterodyne power supply for testing to prevent on-site power frequency electromagnetic radiation and series interference.

>Wide measurement range, up to 5A/5A~25000A/5A or 5A/1A~6300A/1A.

>Has the function of testing current transformer ratio and secondary winding internal resistance.

>Adopts 800×600 high-resolution large-screen color LCD display, with user-friendly interface and operation design, using touch screen auxiliary operation, making operation more convenient and faster.

>Adopts precise software algorithms to further improve the accuracy of measurement data.

>Has intelligent judgment of external wiring status, prompting wiring errors, ratio errors, polarity errors, etc.

>Automatically rounds test data and judges whether it is out of tolerance. Out-of-tolerance data is displayed in orange, and the bottom right corner of the window shows out-of-tolerance, making the data characteristics of the transformer intuitive and clear.

>Directly issues on-site verification conclusions, qualified or out-of-tolerance.

>Large-scale memory can store up to 1000 on-site test data records.

>Uses RS232 or USB interface to connect to computer to print data certificates.

>Uses engineering plastic mold casing, shockproof and pressure-resistant, ensuring the safety of on-site operators and equipment.

Current Transformer On-site Calibration Instrument also has the functions of measuring current transformer error by recursion method and voltage transformer error by potential difference method, facilitating on-site verification of metering devices.

>On-site verification of current transformers does not require standard current transformers, current boosters, load boxes, voltage regulators, and high-current wires. It uses extremely simple test wiring and operation to achieve verification of current transformers,

greatly reducing work intensity and improving work efficiency, facilitating on-site verification of transformers.



Note: The above information and technical parameters are compiled and released by Yangzhou Dary Electric Co., Ltd.





WeChat QR Code
💬
💬 Message ✕

Hello! Have you found what you need? Contact us directly online!

Leave a Message for Consultation
Yangzhou Darui Electric Co., Ltd.
主营产品:Dielectric Loss Tester、Transformer Ratio测试仪、Loop Resistance Tester、高压None线核相仪、Insulation Resistance Tester、Partial Discharge Tester、Cable Fault Tester、DC Resistance Tester等。
☎ Phone: 18952707000(关经理)📞 Landline: 0514-88770377✉ Email: 195214804@qq.com
Verification Code
By submitting, you agree to ourTerms of ServiceandPrivacy Policy