DC High Voltage Generator 120KV 60KV/2mA 5mA 10mA for Qualification Use Detailed Test Operation Explanation
1. Manual Voltage Boost Mode
After power-on, enter the main menu, rotate the setting control lever to move the cursor to the "Manual Voltage Boost" option, click the control lever to enter the manual mode parameter setting interface. After setting various parameters according to specific test requirements, you can enter the voltage boost interface. First press the high voltage on button, after the red light turns on, rotate the knob. When the voltage rises to the set test voltage value, the timer starts timing. After timing is completed, the system will automatically reduce the voltage to zero, then a test record save dialog box will pop up. After making a selection, the test is completed. If the voltage does not reach the set test voltage value, press the 0.75UDC-1mA button, the system will automatically reduce the voltage to 0.75 times the current voltage value, then count down with the set pressurization time. After timing is completed, the system automatically reduces the voltage to zero, then a test record save dialog box pops up. After making a selection, the test is completed.
2. Automatic Voltage Boost Mode
After power-on, enter the main menu, rotate the setting control lever to move the cursor to the "Automatic Voltage Boost" option, click the control lever to enter the automatic mode parameter setting interface. After setting various parameters according to specific test requirements, you can enter the voltage boost interface. At this time, after pressing the high voltage on button, the system will start automatic voltage boost. After rising to the set test voltage value, it stabilizes and starts timing. After timing is completed, the system will automatically reduce the voltage to zero, then a test record save dialog box pops up. After making a selection, the test is completed.
DC High Voltage Generator 120KV 60KV/2mA 5mA 10mA for Qualification Use Product Specifications
|
Specification Technical Parameters |
60/2-10 |
120/2-10 |
200/2-10 |
250/2-5 |
200/300 3/2 |
Other Levels |
|
Output Voltage (kV) |
60 |
120 |
200 |
250 |
200/300 |
600~800kV etc. Custom-made according to contract. |
|
Output Current (mA) |
2-10 |
2-10 |
2-10 |
2-5 |
3/2 |
|
|
Output Power (W) |
120 |
240 |
400 |
500 |
600 |
|
|
Voltage Measurement Error |
<1% |
|||||
|
Current Measurement Error |
<1% |
|||||
|
Ripple Coefficient |
<0.5% |
|||||
|
Working Power Supply |
AC220V±10% |
|||||
|
Ambient Temperature |
-10℃~+40℃ |
|||||
|
Relative Humidity |
≤85% (no condensation) |
|||||